Flame photometer, Sherwood Model 420 Clinical, dual channel emission, with Na, K, and Li filters, fuel and air connections, 1000ppm standard solutions and with lineariser, for 198-264V 50/60Hz a.c., requires a gas regulator and air compressor of the correct type.

FLAME PHOTOMETER SHERWOOD M420 CLINICAL 198-264V 50/60HZ

Double click on above image to view full picture

Zoom Out
Zoom In

More Views

More Information.

Download Product Information

Details

For use in applications examining pharmaceutical and biochemical samples, clinical and biological fluids, industrial processes and environmental testing.

  • Dual channel emission flame photometer with Na, K and Li filters and additionally offering simultaneous Na and K results, built-in lineariser for clinical samples, Li internal standard referencing for enhanced accuracy, microprocessor control and monitoring of measurements, continuous or peak mode operation, and built-in RS232 port for result output.
  • Digital readout with moveable decimal point       
  • Automatic air regulation
  • Auto-ignition of flame                                        
  • Gas cut-off safety feature
  • Integral ‘fail-safe’ flame monitoring                     
  • Recorder output
Model      420 Clinical
Sensitivity             Minimum concentrations producing a display reading of 100 scale units: Na: 0.5ppm. K: 0.5ppm, Li: 2ppm
Detection limts       Na, K <0.02ppm at 2 x standard deviation
Specificity          <0.5% to reading from interferent of Na, K, Li equal to the concentration of element under test
Drift %     <1
Reproducibility    %  <0.5 (Peak ref. mode)
Display channels    2 (LED)
Result display    Auto peak
Result storage   Yes
Outputs   Adjustable analogue and RS232
Dimensions  mm    390 x 345 x 510
Weight kg 10
     
Services required    
Electrical   198-264V 50/60Hz
Air   Moisture and oil free 6 litres/minute at 1kg/cm2
Gas   High grade propane, butane or propane/butane mixture. Fuel should be free from heavy hydrocarbon deposits and regulated to 30 p.s.i. at the cylinder.

 