Cecil System 3 Automatic  Analytical.

High performance, Isocratic modular system for quality assurance, quality control, pharmaceutical testing and analytical separations.

General specifications:

Pump module CE4100 is a dual piston, pulse-free, fully compensated solvent delivery pump with flowrates from 0.001 to 10ml/min. With keypad pump setting and digital control with display and memory storage of up to 30 methods.

Detector module CE4200 is a dual wavelength ratio, UV-VIS detector with range 190 to 700nm with scanning up to 400nm per second and full programming, using up to 6 wavelengths. Keypad operation and 4-line LCD readout with menu guidance, prompts and data display.

The system is controlled using PowerStream System Manager Software CE4900, a Windows-based system control, including programmable wavelength scanning and data processing software operated via the users’ PC. PowerStream is easy to use with a fully interactive graphical interface and display providing full method development.

Supplied with 8µl x 10mm flowcell, column panel, Rheodyne sample injection valve with 20µl loop, solvent reservoir, pipe organiser for 3 solvents and installation kit. All systems are supplied with relevant connecting cables, power supply leads, fuses and operating manual.

Pump specifications:

 Flow rate
 ml per
 minute 
 0.001 to 10 
 Flow reproducibility 
 %
 better than 0.1
 Maximum pressure 
 MPa
 (bar) 
 40
 (400)
 Overpressure cut-out
 MPa
 Adjustable between 3 and 40 
 Low pressure cut-out 
 MPa 
 Adjustable between 0 and 40
 Eluent contact materials 
 
 Biocompatible ceramic, ruby
 and sapphire 
 Dimensions, W x D x H   
 mm
 365 x 290 x 140
 Weight
 kg  7.5
 Power requirements
  115V, 230V 50/60Hz a.c.

 

 

 

 

 

 

 

 

 

 

 

 

 

Detector specifications:

 Wavelength range
   nm  
 190 to 700 
 Wavelength accuracy 
   nm
  ±1
 Wavelength precision 
   nm
 ±0.1
 Bandwidth
   nm
 8 
 Optical bandwidth 
   nm 
 10
 Straylight, at 220nm 
   %
 <0.02
 Absorbance ranges:
 
 0.001 to 2 in 11 ranges
 Noise
   A
 ±0.35 x 10-5 peak to peak
 at 220nm, 2 second constant 
 Dimensions, W x D x H  
   mm
 365 x 290 x 140
 Weight
   kg  17
 Power requirements
  115V, 230V 50/60Hz a.c.