SPECTROPHOTOMETER SP-800-UV 230V 50/60HZ A.C.

SPECTROPHOTOMETER SP-800-UV 230V 50/60HZ A.C.

Product Code: SJ379-15

Manufacturers Code: 83070-06 (formerly Jenway 6850, 6850/230V)

£11,098.49

inc. VAT

£9,248.74

ex. VAT

Download Product Information

Cole-Parmer SP-800-UV

A true, high accuracy, double beam UV/Visible range instrument with variable 0.5, 1, 2, 4 and 5nm spectral bandwidths, using either the integrated controller or remotely by the user’s PC via Windows® compatible Prism PC software supplied.

  • Six measurement modes:
    • — Photometrics and multi-wavelength spectrum scanning (up to 10 wavelengths can be measured, or 20 using Prism PC software, with up to 0.1nm resolution and 2000nm/min. scan speed)
    • — Spectrum scanning (up to 0.1nm resolution and 2000nm/min. speed)
    • — Kinetics (up to 12 hours with time intervals of 0.1, 0.2, 0.5, 1, 2, 5, 10 or 30 seconds)
    • — Quantitation (up to 3 wavelength points using up to 10 standards)
    • — DNA/RNA (ratio, concentration, A320 correction)
    • — Protein analysis (autopeaks and valleys, zoom, addition, subtraction, peak ratios, smoothing, area under curve, wavelength table, derivatives and overlay with PC software)
  • Range: 190 to 1100nm
  • Prism PC software is designed to give a comprehensive range of measurement modes whilst remaining intuitive, user-friendly and providing extensive post-measurement tools and functions
  • Large graphical display
  • Conforms to European Pharmacopoeia standards
  • Range of interchangeable sample chambers allows rapid reconfiguration or upgrade
  • Units with IQ/OQ documentation are available - details on request

Supplied with single 10 x 10mm cuvette holder (sample and reference positions), instruction manual, Prism PC software with USB cable, 2 quartz and 4 glass cuvettes and dust cover. For 230V 50/60Hz single phase supplies.

 Model      SP-800-UV
 Wavelength range x resolution        nm      190 to 1100 x 0.1
 Bandwidths        nm 0.5, 1, 2, 4, 5
 Wavelength accuracy nm ±0.3 (at 0.5 and 1nm), ±0.5 (at 2,4 and 5nm)            
 Straylight       <0.05 (at 220nm and 340nm)
 Source switching range nm 325 to 370
 Photometric range A -0.3 to +3.0
 Photometric accuracy ±0.002
 Photometric reproducibility ±0.001
 Baseline stability A/h ±0.001
 Noise level A 0.0005
 Scan speed nm/minute     3 selectable, from 100 to 200
 Interface   USB and parallel port
 Light source   Tungsten-halogen/deuterium
 Supply requirements   220V 50/60Hz single phase supply
 Overall mm  600 x 450 x 200
 Weight kg  22