- Stainless steel interior and exterior sides (external rear panel-zinc coated steel)
- Temperature range ambient +10˚C to +80˚C
- Fan convection
- With TwinDISPLAY controlCOCKPIT controller
- Heaters located in ribs on all four sides provide uniform heating
- Digital process timer, adjustable between 1 minute and 99 days, 23 hours
- 10 year datalogger
- Setting accuracy: ±0.1˚C
- Supplied with 2 stainless steel wire grid shelves
- Inner glass door
TwinDISPLAY controlCOCKPIT
Display of temperature (˚C or ˚F), fan speed and programme time. Includes two Pt100 DIN Class A temperature sensors in a 4 wire circuit, ethernet and USB interfaces and heatBALANCE function for adjustment of heating power distribution of the upper and lower heating groups in an adjustment range between -50% and +50%. Multiple overtemperature protection monitoring TWW/TWB (protection Class 3.1/3.2, respectively) and mechanical temperature limiter (TB Class 1) to DIN 12 880, with settings for individual maximum/minimum values for over/undertemperature alarms. Also includes AUTOSAFETY function which allows automatic adjustment to the set value within a freely adjustable tolerance range. With AtmoCONTROL logging software for online monitoring of up to 32 connected appliances, optical alarms when set alarm limits are exceeded anda utomatic alarm email notifications. Additionally includes display of the programme sequence created by the user, repeat functions which can be inserted anywhere within the control sequence and creation of repeating weekly programmes which can be transferred via ethernet or USB interfaces provided.
Model | IF260plus | |
Volume | litres | 256 |
Internal (external) | height, mm | 800 (1186) |
width, mm | 640 (824) | |
depth, mm | 500 (684) | |
Shelves supplied | 2 | |
Max. no. of shelves held | 9 | |
Max. loading per shelf grid | kg | 30 |
Max. loading of chamber | kg | 300 |
Power loading | kW | 1.7 |
Power supply | 230V 50/60Hz single phase | |
Weight | kg | 96 |
With TwinDISPLAY controller, integral process timer adjustable from 1 minute to 99 days, 23 hours, multiple overload protection, temperature and time TFT display, chamber ventilation slide adjustment and 2 stainless steel grid shelves. For 230V 50/60Hz single phase supplies.