THERMOSTATIC BATH GRANT TXF200-ST5 5L S/S 230V 50/60HZ

THERMOSTATIC BATH GRANT TXF200-ST5 5L S/S 230V 50/60HZ

Product Code: BJ186-61

Manufacturers Code: TXF200-ST5

£2,665.20

inc. VAT

£2,221.00

ex. VAT

Download Product Information

Grant Optima series thermostatic baths. Complete units including plastic or stainless steel tank, bridge mounting plate and controller.

Optima thermostatic controllers are suitable for most heating tasks in laboratories. Sophisticated electronics and the latest materials combine to ensure optimum performance, accuracy and stability.

Model TXF200 features:

  • PID control, soft-touch rotary selector and push-button setting and on-board programming facility via the control panel, including temperature ramping rate to target temperature
  • Full colour QVGA TFT screen displaying actual and set temperatures, pump speed and memory icons for programming with choice of operating languages (English, French, German, Spanish and Italian).  
  • 99 hours 59 minute timer function and heater control fault indicator, audible and visual alarms
  • USB interface allowing remote control via an external PC using accessory Labwise™ software
  • Programmable relay for ancillary equipment switching during a procedure (external cooling or remote alarm for example)
  • High/low temperature alarm setting and temperature limiting for selected liquid type
  • Memory storage of 10 programmes each of up to 100 segments.
  • User-adjustable overtemperature cut-out
  • Low liquid protection by float switch

With an injection moulded case, heater, variable flow rate external circulation pump and two connector plates with outlets for 6mm or 11mm tubing bores, stirrer, Pt1000 temperature sensor projecting down into the liquid and a socket for an accessory external Pt1000 temperature  probe. For 230V 50/60Hz single phase supplies. Each controller is supplied without clamp. An accessory clamp (BJ180-95) allows the controller to be used with almost any type of vessel where it can be attached to the vessel wall. When the controller is to be used with Optima bath tanks the clamp is unnecessary as a bridge plate is provided

General specifications:

 Model
  
 T100
 TC120  
 TX150
 TXF200               
 Temperature ranges* 
                   without tank
 
                 using S tanks
                                   
                
                 using P tanks
 
 °C
 
 °C
 
 
 °C
 
 0 to 100
 
 0 to 100
 
 
 amb. +5 to 99
 
-20 to 120
 
 0 to 120 (5, 12 and 18L baths)
 -15 to 120 (26 and 38L baths)
 
 amb. +15 to 99 (P5 only)
 otherwise amb. +5 to 99
 
-50 to 150
 
 0 to 120 (5, 12 and 18L baths)
 -15 to 120 (26 and 38L baths)
 
 amb. +15 to 99 (P5 only)
 otherwise amb. +5 to 99
 
-50 to 200
 
 0 to 120 (5, 12 and 18L baths)
 -15 to 120 (26 and 38L baths)
 
 amb. +15 to 99 (P5 only)
 otherwise amb. +5 to 99
 Stability, S tanks
 (DIN 58966) at 70°C
 ±°C
 0.05
 0.05
 0.01        
 0.01
 Uniformity,
 S tanks at 70°C
 ±°C
 0.1
 0.1
 0.05 
 0.05
 Setting resolution 
 °C
 0.1
 0.1
 0.1
(0.01 using labwise software)
 0.1
(0.01 using labwise software)
 Display
 
 4 digit LED
 4 digit LED
 Full colour QVGA TFT
 Full colour QVGA TFT
 Display resolution   °C  0.1  0.1  0.01   0.01
 No. of stored
 temperature values
 
 3
 3
 3
 3
 Two point re-calibration     yes  yes  yes  yes
 Offset adjustment    no   no  yes  yes
 External probe socket  
 
 no
 no
 yes
 yes
 USB interface
 
 no
 no
 yes
 yes
 Programmable
 
 no
 no
 yes, via PC
 yes, via PC and directly
 Number of stored
 programmes
 
 none
 none
 1 x 30 segment
 10 x 100 segment
 Relays
 
 none
 none
 1
 1
 Heater power, 220-240V
 kW
 1.3
 1.3
 1.9
 1.9
 Pump (water) 
             max. pressure
                     max. flow
 
 mbar
 l/min.
 
 -
 -
 
 210
 16
 
 310
 18
 
 530
 23 (adjustable flow rate)
 Inlet/outlet pipe bore
 diameters
 mm
 -
 6/11
 6/11
 6/11
 Overall W x D x H
 excluding clamp
 mm
 115 x 145 x 315
 115 x 145 x 315
 115 x 145 x 315
 115 x 145 x 315
 Height above tank rim
 mm
 200
 200
 200
 200
 Depth below tank rim
 mm
 135
 135
 135
 135

 

 

 

 


 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

*A lid or layer of polypropylene spheres must be used when operating between 60°C and 100°C, and is recommended for use below room temperature. Above 100°C, a lid must always be used. Minimum operating temperature without accessory cooling is ambient plus 5°C, except in the 5 litre tanks, in which it is ambient plus 15°C. Minimum and maximum temperatures achievable are dependent upon the tank insulation and the minimum operating temperature depends on the accessory immersion cooler selected.

ST series tanks

Stainless steel tanks with outer cases of glass-reinforced plastic which has excellent insulating properties and is resistant to acids and laboratory solvents. Operating  temperature  ranges  are  as  indicated.  Supplied  with  bridge  mounting  plate  for  Optima  T-series  controllers  and  configured  to  allow  dual- positioning, either facing along the length or away from the width of the bath, for convenient access. Also include a recessed drain tap (except 5 litre size).

General specifications:

 Reference
  
 ST5
 ST12  
 ST18
 ST26                  
 ST38                  
 Operating range
 °C
 0 to 200
 0 to 200 
 0 to 200 
 -15* to 200
 -15* to 200
 Suitable immersion cooler 
 
 C1G
 C1G
 C1G
 C1G, C2G
 C1G, C2G
 Tank capacity
 litres 
 5
 12
 18
 26
 38
 Top opening
 mm
 150 x 150
 205 x 300
 385 x 300        
 385 x 300
 575 x 300
 Liquid depth, minimum
 mm
 85
 85
 75
 125
 125
 Liquid depth, maximum 
 mm
 140
 140
 130
 180
 180
 Inner tank L x W x H
 mm
 300 x 150 x 150 
 325 x 300 x 150 
 505 x 300 x 150 
 505 x 300 x 200
 690 x 300 x 200
 Outer case L x W x H
 mm
 330 x 180 x 180
 360 x 330 x 180
 540 x 330 x 230
 540 x 330 x 230
 730 x 333 x 230
 Overall L x W x H
 (including controller) 
 mm  330 x 180 x 395  360 x 330 x 395  540 x 330 x 395   540 x 330 x 405  730 x 333 x 405

 

 

 

 

 

 

 

 

 

 

 

  *Using C2G immersion cooler, BJ240-52.