Spectrophotometer Jenway 7310, scanning, Visible, range 320 to 1000nm, 5nm bandwidth, 1nm, 2nm and 5nm scanning intervals, absorbance, concentration, factor and transmission modes, with 10mm square cuvette holder, USB, analogue and RS232 output ports, software CD-ROM, USB memory stick, dust cover, user manual and one hundred x 10mm square, disposable plastic cuvettes, for 100-240V 50/60Hz a.c. 731 001

SPECTROPHOTOMETER JENWAY 7310 VISIBLE SCANNING 100-240V

Double click on above image to view full picture

Zoom Out
Zoom In

More Views

More Information.

Download Product Information

Details

  • Scanning visible, range 320 to 1000nm, Quantitation and Kinetics modes
  • 5nm spectral bandwidth
  • GLP compliant, with logging of time, date and user i.d., instrument and method locks, 240 method memory storage
  • Sealed control panel and display built into hinged lid to avoid splashes and spillages when sample loading
  • Analogue, RS232  and front positioned USB outputs provided
  • Large, graphics display with icon-driven, (non-textual) operation using soft-key navigation
  • Compact design minimises bench space required
  • Accessory printer-module available

Supplied with 10mm cell holder, 100 x 10mm disposable cuvettes, downloading/analysis software CD-ROM, interface cable, 2GB USB memory stick and instructions. With universal power supply for 100-240V 50/60Hz single phase supplies.

 Model      7310
 Wavelength range        nm     320 to 1000
 Bandwidth          nm 5
 Wavelength accuracy        nm  ±2                                 
 Light source         Tungsten-halogen
 Repeatability nm ±0.5 
 Photometric ranges Transmittance         0 to 199.9% x 0.1% resolution ±1% accuracy           
  Absorbance -0.300 to 2.500A x 1A resolution ±0.1 accuracy (at 1.00)             
 Concentration modes range  -300 to 9999
  resolution Selectable 1/0.1/0.01/0.001
  calibration Blank and 1 factor/standard
  factor  1 to 10,000
  standard  1 to 1000
 Quantitation modes range -300 to 9999
  resolution  Selectable 1/0.1/0.01/0.001
  calibration Blank and up to 6 standards
  curve fit Quadratic, linear (and through zero), interpolate
 Kinetics modes time, secs 2 to 9999
  calibration Blank and 1 factor/standard
  resolution  Selectable 1/0.1/0.01/1
  analysis conc. rate of change, initial and final abs. or % trans.
 Scanning modes interval, nm Selectable 1/2/5
  analysis ABS, %T, peak and valleys
 Outputs   Analogue, RS232, USB
 Dimensions H x W x D mm 220 x 275 x 400
 Weight kg 6